Rockwell hardness testing for plastics measures a plastic specimen's resistance to indentation under a defined indenter and load. The result is a dimensionless hardness number on an established scale. Because thermoplastics are viscoelastic and time-dependent, the test must run under strictly controlled conditions to give comparable results. This article covers the measurement principle, common scales, specimen preparation, the operating procedure with HRM versus HRR guidance, result expression and repeatability, suited materials and applications, and the governing standards ASTM D785 and ISO 2039.
Plastic
Rockwell hardness testing — principle and scales
A Rockwell test on plastics presses a hardened steel ball indenter into a flat specimen surface. It starts with a preliminary (minor) load, then applies a total (major) load for a fixed dwell time. The major load is then removed and the preliminary load restored. The permanent increase in penetration depth between the two minor-load readings is converted to a hardness number. For plastics, higher numbers mean greater indentation resistance. Unlike metals, plastics creep under load, so dwell time is part of the definition of the result. The alpha hardness value in ISO 2039-2 derives from the depth difference measured 15 s after load release, expressed on the Rockwell alpha scale. Scales are defined by ball diameter and total load. For plastics, the common ones are R (12.70 mm ball, 60 kgf) and M (6.35 mm ball, 100 kgf). Others, such as L, E, and K, serve softer or harder materials.
Sample
preparation and specimen requirements
Specimens must have two flat, parallel faces free of sink marks, flash, scratches, and dust. Thickness matters because plastics transmit indentation strain. A minimum thickness of about 6 mm is customary for ball scales, or several millimeters when measured away from anvil effects. Thin sheet may be stacked only when layers do not slip and the surfaces are flat. Laminated buildup can also alter creep behaviour. Test faces should sit far enough from edges and from previous indentations so that surrounding material does not distort the reading. Condition specimens to the standard laboratory atmosphere, typically 23 °C with controlled relative humidity, until equilibrium is reached — commonly at least 40 hours for hygroscopic materials. Test moulded specimens on the surface representative of the product, and note any machined surface, since machining can work-harden or heat-affect the skin layer and shift readings.
Test
procedure and scale selection (HRM/HRR)
Place the specimen on a clean, flat anvil and apply the preliminary load without shock. Bring the total load into contact smoothly. Hold it for the specified dwell — commonly 10 to 15 seconds for plastics — then read the dial or digital display while the minor load is still applied, and release the major load. Take readings at multiple points spaced apart, discarding any that show anomalous behaviour. Scale selection follows the anticipated hardness. HRR, with a 12.70 mm ball and 60 kgf total load, suits softer thermoplastics such as polyethylene, polypropylene, PVC, and ABS, where a small ball would sink excessively. HRM, with a 6.35 mm ball and 100 kgf total load, suits harder, more rigid materials such as polycarbonate, nylon, acetal, and phenolic mouldings. A practical rule: choose the scale on which the specimen reads within the useful mid-range of the dial. This ensures the ball neither bottoms out on the anvil through the specimen nor produces an off-scale high value with negligible penetration.
Result expression and repeatability
Report the hardness as a scale designation followed by the numeric value, for example HRR 105 or HRM 80. Include the test standard, dwell time, specimen thickness, and conditioning environment, because these variables influence the number. Values from different scales are not interchangeable; a conversion between HRR and HRM is only an approximation and must not replace a direct measurement on the chosen scale. For a valid determination, take a set of readings — five is a common minimum — and report the arithmetic mean and, where required, the range or standard deviation. Repeatability depends on uniform specimen thickness, flatness, temperature control, dwell-time consistency, and freedom from surface defects. Viscoelastic recovery between readings, anvil effects on thin specimens, and vibration are frequent sources of scatter. If successive readings at one laboratory drift systematically, verify indenter condition, load calibration, and specimen conditioning before attributing the change to the material.
Applicable types and applications
The method applies to many sample types. These include rigid and semi-rigid thermoplastics, thermosetting moulding compounds, and filled or reinforced grades whose surfaces are homogeneous at the scale of the indentation. It is less suitable for very soft elastomers, cellular plastics, and thin films. For those, other instruments give more meaningful data: durometers (Shore A/D) for soft materials, or microindentation for thin sections. Typical uses include incoming inspection of moulding compounds, batch-to-batch quality control, comparison of annealed versus as-moulded parts, screening of formulation changes, and specification conformance in trade documents. Because Rockwell values for plastics are sensitive to time and temperature, they serve best as comparative indices within a defined procedure rather than as absolute design properties. When a load-bearing judgement is needed, stress–strain or creep data should accompany the hardness result.
Related standards (ASTM D785, ISO 2039)
ASTM D785 defines Rockwell hardness of plastics and electrical insulating materials using the ball-indenter scales R, L, M, E, and K, with specified loads, dwell times, and thickness requirements. ISO 2039 exists in two parts. ISO 2039-1 specifies a ball indentation method that reports hardness in newtons per square millimetre from force divided by the projected contact area. ISO 2039-2 specifies the Rockwell hardness method corresponding in principle to the ASTM procedure. Results obtained under the two documents are not automatically equivalent, because load application, dwell definitions, and depth-measurement timing differ. Laboratories must state which standard governed the test and keep the procedure fixed within a comparison programme. When both standards are cited in a specification, verify the scale, dwell time, and specimen thickness clauses befOre testing, and treat any cross-standard conversion as indicative only.
Frequently Asked Questions
What
sample requirements apply to plastic Rockwell hardness testing?
Specimens must have smooth, parallel upper and lower surfaces, sufficient thickness to prevent anvil effects, and adequate area to position indentations away from edges. Submitted samples should be representative, clean, and free of surface defects that could distort the plastic Rockwell hardness result.
How do I choose between methods when plastic Rockwell hardness testing differs from other hardness approaches?
Plastic Rockwell hardness testing measures indentation depth under a defined preload and major load, making it suited to harder or more rigid plastics. Softer, more elastic materials may be better served by ball indentation methods. Selection depends on the material's rigidity and the applicable scale, such as HRM or HRR.
What
criteria are used to judge plastic Rockwell hardness results against reference limits?
Judgments rely on the reported scale designation together with the numeric hardness value, evaluated against the material specification or product standard limits. Because results depend on scale selection and specimen preparation, comparisons are valid only when test conditions and repeatability requirements match those references.